Venturi Sampler w/ air and wash (Single Stream), for BioTector B7000/B7000i/B7000i Dairy

Venturi Sampler w/ air and wash (Single Stream), for BioTector B7000/B7000i/B7000i Dairy
Product #: 19-BAS-001
Unit Price Contact Hach

Covers Long Distances with Multiple Channels

The Venturi Vacuum Sampler is a pump capable of drawing samples from a horizontal distance of up to 80 meters and a vertical distance of up to 7 meters. It can be equipped with 1, 2, 3, 4, or 6 channels allowing you to make upstream adjustments to a downstream analyser and requires pressurized air and hot water connections for operation.

Boasts a Simple, Worry-Free Design

The Venturi Vaccum Sampler creates a vaccum generated by a flow of pressurized air to draw sample content to the BioTector TOC analyser. The Venturi Sampler is powered by and integrated next to the TOC analyser, allowing for ease of installation and access. Unlike mechanical sample pumps, the Venturi Vacuum Sampler requires infrequent maintanance and attention, even with highly charged wastewater containing large amounts of things like oil, grease, or solids.

Requires Little Maintenance While Ensuring Sample Quality

After each measurement cycle the sample is pushed out of the Venturi Vacuum Sampler and associated sampling lines and all components that come in contact with the sample are backwashed with a hot water microbubble mixture. These components are then air dried and are ready to receive the next sample, avoiding any potential cross contamination or dilution and providing a truly representative sample for analysis.

Handles Low Flow with Ease

Where mechanical pumps would be in danger of damages by running dry, the Venturi Vacuum Sampler can be adjusted to low flow rates in the respective sampling channel without the need of a barrier system and comes with an automatic no sample fill fault indication.

What's included?

Sampler only.
Hoses need to be ordered separately.

Specifications

Ambient Temperature: 5 - 40 °C
Dimensions (H x W x D): 2010 mm x 390 mm x 160 mm
Housing Material: PLC Enclosure: Polystyrene with glazed polycarbonate door
Instrument air: Instrument air: 6 bar
(50 L/min flow rate when venturi pump is running at 6 bar)
(~100 L/min flow rate when venturi sampler is purging at 3.5 bar)
Air quality: -20 °C dew point (free of water, oil and dust)
Material: Sample chamber: PFA
Fittings and valves: SS-316, EPDM, PTFE
Venturi pump: SS-316
Sample tubing: PFA (1/2" OD x 3/8" ID)
Number of sample streams: 1
Operating System: PLC with Keyboard
Automatic 'no sample fill' fault indication
Direct control from the BioTector through the 6 core cable
Power Consumption: 24 W (1 Amp 24 V DC), power supplied by BioTector analyser
Requirement rinse water pressure: 3 bar (~10 L/h at 3 bar)
Water can be heated up to 90 °C for hot water backwash
Water regulator and/or water heater to be supplied by the user
Sample Inlet Temperature: 2 - 90 °C
Sample Pressure: Ambient pressure
Sample Quality: Only for non-foaming samples!
Sampling Principle: Vacuum sampling
Service Interval: 6 months (typical)
Suction Height: 7 m (at sea level; measured from the bottom of the sampler to the lowest liquid level)
Suction Hose Length: Max. distance of sample transport: 80 m
Weight: 20 - 30 kg
What's included?: Sampler only.
Hoses need to be ordered separately.