The Method 8282 has specific recommendations for maintenance of the labware and cleaning the Pour-Thru cell.
Labware cleaning:
Fully clean all containers that are used in this test to remove any traces of silica. Use plastic containers for all analysis and storage because glass can contaminate the sample with silica. Small bottles or flasks with screw-type closures work well.
- Clean containers (do not use phosphate detergents), then rinse with high quality deionized water of low-level silica concentration.
- Soak for 10 minutes with a 1:50 dilution of Molybdate 3 Reagent in low-level silica water.
- Rinse repeatedly with either low-level silica water or the sample before use. Keep containers tightly closed when not in use.
- Fill the Pour-Thru Cell with this same mixture of Molybdate 3 and water. Soak for 10 minutes.
- Rinse with low-level silica water.
Keep containers tightly closed when not in use and rinse the labware repeatedly with either low-level silica water or the sample before use.
Pour-Thru cell cleaning:
The Pour-Thru Cell can collect a buildup of products with color, especially if the reacted solutions stay in the cell for long periods of time after measurement.
- Rinse the Pour-Thru Cell with a 1:5 dilution of ammonium hydroxide solution to remove the color.
- Fully rinse with deionized water.
- Put a cover on the Pour-Thru Cell funnel when it is not in use.
The following can also be done to clean the Pour-Thru cell:
- Clean Pour-Thru cell assembly with phosphate free detergent, then rinse with high quality DI water.
- Soak for 10 minutes in 1:50 dilution of molybdate 3 reagent in low-level silica water. This should rid the sample of any adhering silica.
The EZ Series EZ1034 can analyze from 1 µg/L up to 100 µg/L of Si and the EZ1035 can measure up to 1 mg/L Si with an internal dilution.
The reagent 1 (R1) blank value can be found on the reagent bottle. If the sample has very low silica concentrations and has issues with high readings (one to two ppb high), which are not caused by blueing of the sample cell or instrument issues, then a field blank determination can be performed to adjust the blank value. Clean the sample cell by following the procedure listed in the user manual.
The method used to determine the blank on the 5500sc Silica Analyzer is the standard additions method. This method requires a stable sample concentration feed to the analyzer for the duration of the testing, which will be a couple hours. Begin by performing a grab sample test with the sample water, and record the analyzer reading. Collect 500 ml of sample, and to this add 2 ml of 500 ppm standard, then run this spiked sample as a grab sample and record the reading. Collect 500 ml of sample, and to this add 4 ml of 500 ppm standard, then run this spiked sample as a grab sample and record the reading. Repeat running grab samples spiked with 6, 8, and 10 ml, recording each reading.
When the spiked grab sample readings have been collected, then chart these on graph paper (or in Excel). Spike value - 0, 2, 4 etc., ppb is the x-axis; sample reading in the y-axis. Draw a best-fit straight line through the charted points. Where the line crosses the x-axis is the value of the reagent blank. (The intercept will be a negative value, but it is entered as a positive blank value.)
Determination of reagent blank is performed at concentrations near the detection limit of the analyzer, and the spiked sample readings noted above are subject to noise and interference. The labware and mixing vessels must be scrupulously clean for good results. It is best to repeat each standard addition at least three times to assure results are repeatable. Final accuracy of the blank determination will probably be around +/- 1 ppb, and very likely in this range of agreement with the value on the Molybdate reagent bottle.
A disagreement of 1-2 ppb between two analyzers running sample that is itself only 1-2 ppb (or even less) can be very difficult to isolate. If the process system is stable and runs a known baseline response following regeneration, and no other fault can be found in the analyzer, the most practical solution may be to adjust the blank value on this analyzer so that readings match the known baseline.
The reagents and standards for the 5000sc have the following shelf life:
5500sc silica reagent set product # 6783600 includes:
- 5500sc reagent 2 silica 2-liter product # 6774902 has a shelf life of 720 days from the date of manufacture unopened.
- 5500sc standard 1 silica 2-liter product # 6775002 has a shelf life of 360 days from the date of manufacture unopened.
- 5500sc reagent 3 silica 1.8-liter product # 6775202 has a shelf life of 720 days from the date of manufacture unopened.
- 5500sc reagent 3 silica powder 205G product # 6775355 has a shelf life of 1800 days from the date of manufacture unopened unprepared.
- 5500sc reagent 1 silica 2-liter product # 6774802 has a shelf life of 720 days from the date of manufacture unopened.
The exact date and month of expiration will be printed on the label for each bottle.
The reagents for the 5500 silica/phosphate analyzer are a different concentration versus the older Series 5000 silica/phosphate analyzers reagents and therefore, are not compatible. Due to the difference in concentrations, the reagent volumes are different and can cause accuracy issues on a different model of analyzer.
Please note, the calibration standards for the 5500sc silica/phosphate analyzers are the same with the Series 5000 model.
If a standard bottle for the Series 5000 is unopened and has not expired, the contents may be used to calibrate for the same parameter on a 5500sc analyzer.
Before use transfer the content to the correct bottle style for the 5500sc systems. An incorrect bottle type may result in an air or standard leak that could damage the system.