Accuracy:
±2% or 0.01 NTU from 0 - 40 NTU
±10% of reading from 40 - 1000 NTU based on Formazin primary standard
Cable Length:
1.6 m, extendable up to 50 m for instrument without accessories
Calibration Method:
For Formazin and Stablcal:
20 NTU from 0 - 40 NTU; at 20 FNU and 600 NTU for full range
Custom calibration for up to 6-point calibrations
Certifications:
CE compliant
US FDA accession number: 1420493-000 EPA version, 1420492-000 ISO version
Australian ACMA Marking
Communication:
None
Controller:
Sensor Only
Dimensions (H x W x D):
249 mm x 268 mm x 190 mm
Enclosure rating:
Electronic compartment IP55; all other functional units IP65 with process head/ACM attached to the TU5300sc/TU5400sc instrument
Fitting:
Sample quick connector: ¼-in. for ¼-in. tubing
Instrument:
With Automatic Cleaning Module
Light source:
Class 2 laser product, with embedded 650 nm (EPA 0.43 mW) or Class 1 laser product, with embedded 850 nm (ISO), max. 0.55 mW (complies with IEC/EN 60825-1 and to 21 CFR 1040.10 in accordance with Laser Notice No. 50)
Lower Limit of Detection (LOD):
0.002 NTU
Material:
ASA Luran S 777K / RAL7000, TPE RESIN Elastocon STK40,
Thermoplastic Elastomer TPS-SEBS
Model:
TU5300sc
Mounting Configurations:
Wall Mount
Operating Humidity:
Relative humidity: 5 - 95% at different temperatures, non-condensing
Operating temperature range:
0 - 50 °C
Options:
Process Head & Automatic Cleaning Module
Power requirements (Voltage):
None
Range:
EPA:
0 - 700 NTU / FNU / TE/F / FTU
0 - 175 EBC
Regulatory:
EPA
Repeatability:
Better than 1% of reading or ±0.002 NTU on Formazin at 25 °C (77 °F), whichever is greater
Resolution:
0.0001 NTU / FNU / TE/F / FTU / EBC
Response time:
TU5300sc: T90 <45 seconds at 100 mL/min
TU5400sc: T90 <30 seconds at 100 mL/min
Sample Flow Rate:
100 - 1000 mL/min; optimal flow rate: 200 - 500 mL/min
Sample Pressure:
6 bar maximum, compared to air at sample temperature range from 2 - 40 °C
Sample Temperature:
2 - 60 °C
Signal Average Time:
TU5300sc: 30 - 90 seconds
TU5400sc: 1 - 90 seconds
Storage conditions:
-40 - 60 °C
Stray Light:
<10 mNTU
System Check:
No
Units:
NTU, FNU, TE/F, FTU, EBC
Verification:
Liquid: Stablcal, Formazin (0.1 - 40 NTU)
Dry: Glass Rod at <0.1 NTU
Warranty:
12 months
Weight:
2.7 kg (5.0 kg with all accessories)
What's included?:
TU5300sc Turbidimeter, Automatic Cleaning Unit, User Manual, Wall Mount, Vial Replacement Tool, Screw Set, Drying Cartridge, Flow Regulator, Service Bracket