Analogue Outputs:
Five 0-20 mA or 4-20 mA analog outputs on analog output module
Cable Length:
10 m (33 ft)
Calibration Method:
Zero: Electrically or with oxygen free water, Slope: in air or against a laboratory measurement
Communication:
Analog:
Five 0 - 20 mA or 4 - 20 mA analog outputs on each analog output module
Up to two analog Input modules (0 - 20 mA or 4 - 20 mA). Each input module replaces a digital sensor input.
Digital:
Profibus DPV1 module
Modbus TCP
Profinet IO module
Ethernet IP module
Communication: digital:
Modbus RS232/RS485, Profibus DPV1, Profinet optional
Communication Capabilities:
LAN + Profibus DP
Conduit Openings:
½" NPT Conduit
Connection Drain Line:
8 mm tubing
Connection Sample Line:
¼" NPT thread (6 mm or ¼" tubing advised)
Display:
3.5-inch TFT color display with capacitive touchpad
Electrical Certifications:
CE. ETL certified to UL and CSA safety standards (with all sensor types), FCC, ISED, KC, RCM, EAC, UKCA, SABS, CMIM, Morocco
Enclosure rating:
UL50E type 4X, IEC/EN 60529–IP 66, NEMA 250 type 4X
Metal enclosure with a corrosion-resistant finish
Lowest Detection Limit :
< 1 ppb
Maintenance Interval:
Membrane Lifetime: 6 months depending on sample
Material:
Electrodes: Gold cathode, Silver anode; Flow-through cell: Stainless Steel 316L; Membrane: PFA, Sensor Body: Noryl
Material Enclosures:
Polycarbonate Aluminium (powder coated) Stainless Steel
Measuring range:
0 to 2000 ppb (0-2 ppm)
Operating temperature range:
-20 - 60 °C (-4 - 140 °F) (8 W (AC)/9 W (DC) sensor load)
-20 - 45 °C (-4 - 113 °F) (28 W (AC)/20 W (DC) sensor load)
Linear derating between 45 - 60 °C (-1.33 W/°C)
Power requirements (Hz):
50/60 Hz
Power requirements (Voltage):
100 - 240 VAC
Power supply:
Without power cord
Range:
0 - 2000 ppb (0 - 2 ppm)
Relay: Operational mode:
Primary or secondary measurement or timer/scheduler
Relay functions:
Scheduler (Timer), Alarm, Feeder Control, Event Control, Pulse Width Modulation, Frequency Control, and Warning
Relays:
Two relays (SPDT);
Wire gauge: 0.75 - 1.5 mm² (18 - 16 AWG)
AC controller
Maximum switching voltage: 100 - 240 VAC
Maximum switching current: 5 A Resistive/1 A Pilot Duty
Maximum switching power: 1200 VA Resistive/360 VA Pilot Duty
DC controller
Maximum switching voltage: 30 VAC or 42 VDC
Maximum switching current: 4 A Resistive/1 A Pilot Duty
Maximum switching power: 125 W Resistive/28 W Pilot Duty
Repeatability:
±0.5 ppb or ±5%, whichever is greater
Reproducibility:
±0.5 ppb or ±2% whichever is greater
Response Time:
For step change 1 - 40 ppb: < 30 s
Sample Flow Rate:
66 - 166
Security levels:
2
Software available:
Non-Cloud
Storage conditions:
-20 - 70 °C (-4 - 158 °F)
Temperature Compensation:
Automatic in the range of 0 - 45 °C (32 - 113 °F)
Units:
mg/L, ppm, µg/L, ppb, mbar, hPa, inch Hg, mmHg
Warranty:
12 months
Weight:
3.2 kg
What's included?:
Controller, mounting hardware, oxygen sensor, temperature sensor, cable, flow-chamber, basic user manual