Accuracy:
Using standard solutions: 2% ± 0.05 mg/L
Cable Length:
2 m fixed data cable at analyzer
2 m fixed power cable at analyzer
Channels:
1
Data Transfer:
Current output, relays and bus communication via SC controller
Dimensions (H x W x D):
720 mm x 540 mm x 390 mm
Enclosure rating:
IP55
Filtersystem included:
No, use e.g. Filtrax
Flow:
1 - 20 L/h Sample (free of suspended solids)
Fuse Protection:
Via SC1000 controller or power box. Maximum of 2 analysers per SC1000 or power box
Housing Type:
Outdoor
Indoor/Outdoor:
Outdoor
Lower Limit of Detection (LOD):
0.05 mg/L PO4-P
Material:
ASA/PC UV-resistant
Measurement method:
Photometric method using vanadate-molybdate
Measuring Interval:
5 - 120 min, adjustable per 5 min.
Measuring Principle:
Phosphate ions react with vanadate-molybdate reagent to form a yellow dye.
Measuring range:
0.05 - 15.0 mg/L PO₄-P
Number of Channels:
1
Operating temperature range:
-20 - 45 °C
Outputs:
Current output, relays and bus communication via SC controller
Parameter:
ortho-Phosphate
Permissible Chloride range:
Max. Cl- concentration: 1000 mg/L
pH Range:
5 - 9 pH
Power requirements (Hz):
50 Hz
Power requirements (Voltage):
115 - 230 VAC, powered by Power Box or SC Controller
Power supply:
115-230 VAC
Pressure Range:
-30 - 50 mbar with continuous sample preparation; at overflow vessel
Reagent Consumption:
500 mL/month
Reproducibility:
2 % + 0.05 mg/L
Response Time:
< 5 min
Sample Delivery Hose Options:
Outdoor
Sample Level:
Level of liquid in basin with filtration probe must be below analyzer
Sample Quality:
Ultra filtrated or comparable
Sample Temperature:
4 - 40 °C (39 - 104 °F)
Sampling Option:
1 channel continuous sample from filtration unit, e.g. Filtrax
Storage conditions:
-20 - 60 °C (-4 - 140 °F), 95% relative humidity, non-condensing
Warranty:
12 months
Weight:
Without sample preparation system and without chemicals: 31 kg
What's included?:
0.05 - 15 mg/L PO₄-P, 115-230 VAC Phosphax sc Ortho-Phosphate Analyzer, for one channel continuous sample, 3 month supply of all necessary consumables at 5 minute cycle time, user manuals