Analogue Outputs:
3 smart 0/4-20 mA (500 ohms), programmable as linear or tri-liniear, configurable to send diagnostics information
Communication:
3 x 0/4-20 mA; RS485; Ethernet; Profibus
Communication Capabilities:
3 x 0/4-20 mA; RS485; Ethernet
Compatible Instruments:
Channel 1 : 311xx or A1100/GA2400
Channel 2 : 312xx
Compliance Certifications:
EMC: EN61326-1:2006
CE: EN61010-1:2010
ETL, conforming to UL 61010-1 and CSA 22.2 No. 61010-1
Controller:
512, Multi channel Controller EC/TC /LDO w/ inter-channel compensation (Nuclear)
Data storage:
Rolling buffer or store once mode for up to 1000 measurements and 1000 operator actions
Holds calibration records for last 50 calibrations
Dimensions (H x W x D):
156 mm x 220 mm x 253.5 mm
Display:
Colour TFT touchscreen display
Electrode Compatibility:
Channel 1 : 311xx or A1100
Channel 2 : 312xx
Enclosure Construction:
Panel Mount
Material:
Aluminum
Measurements:
Resolution, accuracy and response time are determined by sensor
Mounting:
Panel Mount
Operating conditions:
0 - 95 % non-condensing relative humidity
Operating temperature range:
-5 - 45 °C
Parameter:
O₂ (EC), H₂ (EC)
Password Protection:
5 levels of authorised access to configuration and data management
Power:
Universal 100/240 VAC @ 50/60 Hz, 25 VA
Power requirements (Voltage):
100 - 240 VAC
Power supply:
Universal 100/240 VAC @ 50/60 Hz, 25 VA
Relays:
3 measurement alarm relays (2A to 30 VAC or 0.5 A to 50 VDC); configurable to send diagnostics information
1 system alarm relay (2 A to 30 VAC or 0.5 A to 50 VDC)
Sensor Connectors:
Standard Cable Glands
Units:
• Gas concentration: Configurable for gas or liquid phase with multiple unit options
• Pressure : External and barometric pressure with multiple unit options
• Temperature : Sample temperature with unit options (K, °C, °F)
User Interface:
Touch screen panel displays: concentration, trend graph, diagnostics, alarm status, historical data
Warranty:
12 months
Weight:
2.9 kg
What's included?:
512 controller and manual only.
Sensors and cables, flow-chamber or process connection, must be ordered separately.