Column, Cadmium, for Nitrate determination

Product #: 50237A
Unit Price Contact Hach
Ships within 1 week

 

Lachat Cadmium Columns (Product Number 50237A) are specialized, packed columns used on the Lachat Flow Injection Analyzer to reduce nitrate to nitrite for water quality testing.

Key Specifications & Maintenance

Function
Reduces nitrate (NO₃2-) to nitrite (NO₃2-) so it can be colorimetrically analyzed using the Greiss reagent. The Greiss reagent works via a diazotization reaction with nitrite to form a highly colored, pink/purple azo dye.

Efficiency
Calculated by dividing the nitrate response by the nitrite response:
E= (NO₃2-/ NO₃2-) x 100

Lifespan Tip
Adding 2 drops of 2g/100 mL copper sulfate solution per liter of buffer can extend column life by keeping the cadmium properly copperized.

  Hazardous Items

Items with this mark may be considered hazardous under some shipping conditions.

If necessary, we will change your selected shipping method to accomodate these items.