Accuracy:
±2% of reading or ±0.02 NTU (whichever is greater) from 0 - 40 NTU
Accuracy 2:
±5% of reading from 40 - 100 NTU (Defined according to ISO 15839.)
Alarm:
When connected to sc controller: three set-point alarms, each equipped with an SPDT relay with unpowered contacts rated 5A resistive load at 230 Vac
Certifications:
CE certified by Hach Company to EN61326 (industrial levels)
Listed by ETL to UL 61010A-1: Certified by ETL to CSA C22.2 No. 1010.1: CE certified by Hach Company to EN 61010-1
Communication: digital:
When connected to sc controller: MODBUS RS232/RS485, PROFIBUS, HART
Compliance Certifications:
Standard Methods 2130B, USEPA 180.1, Hach Method 8195
Controller:
Single Digital
Dimensions (H x W x D):
254 mm x 305 mm x 406 mm
Measuring range:
0.001 - 100 Nephelometric Turbidity Units (NTU)
Mounting:
Turbidimeter body and head assembly: wall and floor stand sc controller: wall, pole, panel, and floor stand
Number of Channels:
1
Operating Humidity:
5 - 95 % non-condensing
Operating temperature range:
0 - 40 °C (Double Sensor System)
Operating Temperature Range:
0 - 50 °C (Single Sensor System)
Outputs:
Up to five selectable for 0-20 mA or 4-20 mA; output span programmable over any portion of the 0-100 NTU range; built into the sc Controller
Power requirements (Hz):
50 - 60 Hz
Power requirements (Voltage):
115 - 230 VAC
Repeatability:
Better than ± 1.0 % of reading or ±0.002 NTU, whichever is greater (Defined according to ISO 15839.)
Response time:
Initial response in 1 minute, 15 seconds for a full-scale step change
Sample Flow Rate:
200 - 750 mL/minute (3.1 - 11.9 gal/hour)
Sample Input:
1/4 inch NPT female, 1/4 inch compression fitting (provided)
Sample Output:
1/2 inch NPT female, 1/2 inch hose barb (provided)
Sample Temperature:
0 - 70°C (32 - 122°F)
Ship Wt. (lbs):
1720E Turbidimeter: 5.9 kg (13 lbs.)
Signal Average Time:
User selectable from 0, 6, 30, 60, 90 seconds
Storage conditions:
-20 to 60 °C (-4 to 140 °F)
Warranty:
12 months
What's included?:
Hach SC200 Controller, Mounting Hardware, SC controller Basic User Manual, 1720E, 1720E User Manual, and 1000 mL 20 NTU Stablcal . (The controller power cord is ordered separately.)